AbhilekhBlood banks & transfusion · 2–6 °C · ≤ −30 °C · 20–24 °C

Blood bank and blood component storage monitoring

A blood bank runs three different temperature bands in one room, and every one of them has a separate legal limit, a separate failure mode, and the same consequence if it slips.

Blood component storage is unusual: a single department maintains whole blood and red cells at 2–6 °C, fresh frozen plasma at −30 °C or below, and platelets at 20–24 °C under continuous agitation. Three bands, three sets of equipment, three alarm philosophies — and in most facilities, three different logbooks.

The clinical stakes remove any tolerance for a gap in the record. A refrigerator that drifts to 8 °C shortens red-cell viability; a platelet agitator that stops is a contamination risk within hours; a plasma freezer that reaches −20 °C compromises labile clotting factors. Each of these is a discard decision that has to be defensible from the record alone, often months later.

Abhilekh puts all three bands into one fleet view. The same instruments and the same audit chain cover the refrigerator, the freezer and the agitator room, so the department has one system to validate, one calibration programme to run, and one report to hand an inspector.

What you must be able to prove

The expectations behind the record

What an inspector or auditor looks for in this use case, and the framework each expectation comes from.

Drugs Rules Sch. F Part XIIB

Storage bands defined per component, monitored continuously

Whole blood and red cells at 2–6 °C, fresh frozen plasma at −30 °C or below, and platelets at 20–24 °C with continuous gentle agitation. Each requires continuous temperature monitoring with an audible alarm and a recorded history.

NBTC standards

Alarms that alert before the product is compromised

Alarm set points are placed inside the storage limits so staff can respond before the band is actually breached, with escalation to someone who can act out of hours.

AABB · WHO

A 24-hour recorder independent of the unit's own controller

The controller that holds the temperature should not be the only thing that reports it. Independent monitoring is what turns a control loop into evidence.

21 CFR §11.10(c)

Records retained and retrievable for the statutory period

Monitoring records, alarm events and the assessment of each excursion are retained with the donation and issue records they relate to, retrievable in readable form for the retention period.

Why it goes wrong

The failure modes nobody budgets for

Every one of these is a real reason monitoring programmes lose their audit trail — or lose the trust of the people who work with them.

01

Three bands, three logbooks, one inspection

Separate charts for the fridge, the freezer and the agitator make cross-checking a manual exercise, and the one that gets skipped is always the one the inspector picks.

02

Chart recorders that run out of paper on a Sunday

Circular chart recorders are still common and still fail silently. A missing chart week is an unprovable week.

03

Agitator failure that temperature alone does not reveal

A platelet agitator can stop while the room stays at 22 °C. Temperature monitoring passes; the product does not.

04

Out-of-hours response that depends on who is on shift

The alarm sounds in an empty room at 02:00. Without escalation to a phone, the first person to know is the technician who arrives at 08:00.

How Abhilekh covers it

Engineered for this, specifically

Each point below maps to a documented platform behaviour — not a marketing promise.

All three bands in one fleet view

ABH-200-E and ABH-200 cover 2–6 °C refrigerators with buffered or cabled probe measurement, ABH-100-D handles plasma freezers with the logger outside and a Class-A PT100 inside, and ABH-200-H covers the 20–24 °C platelet room with humidity alongside temperature.

Escalation that reaches a human out of hours

Notifications route by email, SMS and push (iOS push in progress) with configurable escalation recipients and role-gated escalation, and the mobile app lets the on-call technician acknowledge and annotate from home with biometric login.

Independent of the unit's own controller

Abhilekh loggers measure with their own calibrated probes on their own timebase, so the record is independent evidence rather than a copy of what the cabinet's controller believes.

Every event lands in one audit chain

Readings, alarms, acknowledgements, assessments and e-signatures enter a single SHA-256 checksum chain with identity, timestamp and meaning of signature — across all three bands and every cabinet.

Nothing lost when the network drops

Each logger holds a CRC-checked ring of up to 409,600 readings plus a never-erased audit region, and backfills into the same chain when connectivity returns.

Recommended hardware

The ABH loggers for this job

Abhilekh runs its own logger line end to end — one manufacturer owns the chain from probe accuracy to the signed record.

Probes for this application
  • ▸PRB-PT100-3W — Class-A immersion probe for freezers and buffered fridge duty
  • ▸PRB-SHT45-1M5 — cabled T+RH head for through-the-seal refrigerator monitoring

The ABH line is at the design-frozen, pre-production stage. Every battery-life, drop, thermal and IP figure is a calculated or simulated design estimate, and the regulatory marks referenced on this site are a certification roadmap — not held certificates.

FAQ

Questions, answered

Yes, and that is the point. ABH-200-E or ABH-200 covers 2–6 °C refrigerators, ABH-100-D covers plasma freezers at −30 °C and below with the logger outside and a Class-A PT100 probe inside, and ABH-200-H covers the 20–24 °C platelet room with humidity alongside temperature. One platform, one validation exercise, one calibration programme, one report.
It replaces what they were there to do — a continuous, independent, retrievable record — without the paper, the ink, and the week that goes missing when a chart runs out. Whether you remove the existing recorders is a decision for your quality system and your local regulator; many blood banks run both during the validation period and then retire the paper.
Alerts route by email, SMS and push, with configurable escalation recipients and role-gated manual escalation (automatic time-based escalation is on the roadmap). The mobile app supports biometric login so an on-call technician can acknowledge, annotate and start the assessment from home — and that acknowledgement, with its timestamp and identity, enters the same audit trail as the reading that triggered it.
Temperature alone will not tell you that, and we will not claim it does. ABH-300-SH adds an accelerometer channel that records motion alongside the environmental record, which is the honest way to evidence agitation. Talk to us about how that maps onto your specific agitator model before you specify it.